Optical oxygen transducer for permanent measurements in salt water
HIGHLIGHTS:
GENERAL:
The oxygen transmitter GODOX 200 is a robust measuring system for low-maintenance continuous operation. Compared to electrochemical sensors, it does not need any electrolyte. Measurement is done with a fluorescence transit time method. Together with the complete data processing, including automatic ambient pressure and temperature compensation, it represents a carefree package for continuous measurements. Measurement at depths of up to 30 m is possible. The lifetime of the replaceable measuring diaphragm is usually 2 years.
TECHNICAL DATA:
Measuring ranges (both can be used simultaneously)
Oxygen concentration:
0 ... 20 mg/l (=ppm)
Oxygen saturation:
0 ... 200 % O2
Output signal:
4 ... 20 mA or 0 ... 5 V (convertible)
Accuracy:
±0.1 mg/l below 1 mg/l, ±0.2 mg/l above 1 mg/l
Response time T90:
< 30 s d
Operating temperature:
0 ... 65 ºC
Supply:
5 ... 15 V DC, approx. 160 mA
Material:
Housing:
PVC/ stainless steel, option -PS "seawater resistant": PVC
Diaphragm:
PET
Dimensions Sensor:
Length:
225 mm
Insertion length:
70.5 mm
Diameter:
42.1 mm
Installation diameter:
28.0 mm
Process connection:
1" NPT front/ rear (others on request)
Connection:
Loose cable ends:
No Color Description
1 red Supply +
2 black Supply -
3 green Output O2 Concentration
4 white Output O2 Saturation
Cable length:
5 m
Other cable lengths on request.
Scope of delivery:
transmitter (consisting of sensor body and evaluation electronics, connected by cable), storage cap