atmospheric oxygen sensor, closed sensor type
Universal application, diving gas, longlife, suitable for GMH 569x
GENERAL:
APPLICATION:
Suitable for measuring in normal atmosphere and in systems without or with slight under or over pressure. The sensor type features a screw thread and can be built in gas-tight in almost every system directly resp. with tube-adapter.
SPECIFICATIONS:
Sensor element:
GOEL 370
Application:
Standard for diving gas in recreational diving, protective gases with increased CO2 content, and more.
Measuring range
O2 partial pressure:
0 ... 350 hPa
O2 concentration:
0.0 ... 100.0 vol. % O2, recommended for 0.2 ... 35.0% by volume O2.
Response time:
90% after approx. 10 s
Accuracy:
< 35 vol.% O2: - 0.2 ... +0.35 vol.% O2.
35-100 vol.-% O2: not specified
Scope of supply:
sensor, flow diverter, T-piece