atmospheric oxygen sensor, closed sensor type with pressure connection
Universal application, diving gas, longlife, with pressure connection, suitable for GMH 569x
GENERAL:
For devices with external pressure port (GMH 5695 / 3695) is this housing optimal. Especially for systems with high or low pressure or with existing backpressure by the flow.
APPLICATIONS:
It can be screwed airtight (Attention: Observe permissible operating pressure!). The device-pressure port is connected to the sensor pressure port. The device measures and compensates for the actual pressure at the sensor.
SPECIFICATIONS:
Sensor element:
GOEL 370
Applications:
Standard for diving gas in recreational diving, protective gases with increased CO2 content, and more.
Measuring range
O2 partial pressure:
0 ... 350 hPa
O2 concentration:
0.0 ... 100.0 vol % O2, recommended for 0.2 ... 35.0% by volume O2
Response time:
90% after approx. 10 s
Accuracy:
< 35 vol.% O2,: - 0.2 ... +0.35 vol.% O2
35-100 vol.-% O2: not specified
Scope of supply:
sensor, flow diverter, T-piece